Sensor specifications |
Analyze | Gaseous or dissolved oxygen |
Measurement range | 1-6000 ppm,
0.1-250 µg/l,
0.1-250 ppb |
Resolution | 0.1ppb, 0.1µg/l (dissolved oxygen)
3ppm (gaseous) |
Detection limit | 0,01ppb |
Accuracy | +/- 1% upper range value |
Response time (t90) | < 60 s |
Operating conditions |
Allowable temperature range during measurement on the tip | -4 to +50 °C |
Storage temperature | -20 to +60 °C |
Cleaning | CIP / disinfectable
5% NaOH, 90°C, 30min
3% H3PO4, 50°C, 30min |
Protection class | waterproof according to IP67 / NEMA 6 |
Design characteristics |
Temperature compensation | automatically by integrated temperature sensor
5 - 50 °C
< 2 % |
Pressure compensation | automatically via input
500 - 1100 hPa |
Salinity correction | automatically via input
0,01 - 70 SAL |
Length of cable | Standard: 1,5; customer specific length available |
Sensor parts touched by media | Stainless steel (316L/1.4435), Sapphire, Silicone (FDA) |
Seals touched by media | Silicon (FDA) |
Probe diameter | 12 mm |
Probe length | several length are available |
Connection | PG13,5 or Swaglock directly onto the shaft |
Data logging | Internal data logger |
Interfaces | RS485 with Modbus RTU ,2x 4(0)-20mA (or 2x 0-5VDC) with Watchdog-Function, Trigger-input and alarm-output channels;
Optional: USB Converter with application software |
Power supply |
 | 12VDC - 36VDC |
 |  |